Proline Promass F 500 EtherNet/IP
System integration
111
Description
Byte
13. Oil density unit
73-74
14. Water density unit
75-76
1)
Only available with the Petroleum application package
Heartbeat monitoring fixed input assembly (Assem112), 100 byte
Description
Byte
1. Mass flow fixed input assembly
1-32
2. Signal asymmetry
33-36
3. Oscillation frequency 0
37-40
4. Oscillation frequency 1
41-44
5. Oscillation amplitude 0
45-48
6. Oscillation amplitude 1
49-52
7. Oscillation damping 0
53-56
8. Oscillation damping 1
57-60
9. Tube damping fluctuation 0
61-64
10. Tube damping fluctuation 1
65-68
11. Exciter current 0
69-72
12. Exciter current 1
73-76
13. HBSI
77-80
14. Frequency fluctuation 0
81-84
15. Frequency fluctuation 1
85-88
16. Electronic temperature
89-92
17. Carrier pipe temperature
93-96
18. Verification status
97-98
19. Verification results
99-100
1)
Only available with the Heartbeat Verification application package.
Configurable input group
Input assembly custom (Assem101), 88 byte
Description
Format
1. - 10. Input values 1 to 10
Real
11. - 20. Input values 11 to 20
Double integer