Electrical connection
Proline Promass F 500 EtherNet/IP
40
2
3
C
1
3
B
1
3
A
1
3
A
1
3
B
2
3
C
4
5
6
A0032476
1
Proline 500 digital transmitter
2
Proline 500 transmitter
3
Promass sensor
4
Non-hazardous area
5
Hazardous area: Zone 2; Class I, Division 2
6
Hazardous area: Zone 1; Class I, Division 1
A
Standard cable to 500 digital transmitter → 40
Transmitter installed in the non-hazardous area or hazardous area: Zone 2; Class I, Division 2 / sensor
installed in the hazardous area: Zone 2; Class I, Division 2
B
Standard cable to 500 digital transmitter → 41
Transmitter installed in the hazardous area: Zone 2; Class I, Division 2 / sensor installed in the hazardous
area: Zone 1; Class I, Division 1
C
Signal cable to 500 transmitter → 43
Transmitter and sensor installed in the hazardous area: Zone 2; Class I, Division 2 oder Zone 1;
Class I, Division 1
A: Connecting cable between sensor and transmitter: Proline 500 – digital
Standard cable
A standard cable with the following specifications can be used as the connecting cable.
Design
4 cores (2 pairs); uninsulated stranded CU wires; pair-stranded with common
shield
Shielding
Tin-plated copper-braid, optical cover ≥ 85 %
Loop resistance
Power supply line (+, –): maximum 10 Ω
Cable length
Maximum 300 m (1
000 ft), see the following table.
Cross-section
Cable length [max.]
0.34 mm
2
(AWG 22)
80 m (270 ft)
0.50 mm
2
(AWG 20)
120 m (400 ft)
0.75 mm
2
(AWG 18)
180 m (600 ft)