Proline Promass F 500 EtherNet/IP
Electrical connection
43
C: Connecting cable between sensor and transmitter: Proline 500
Standard cable
6 × 0.38 mm
2
PVC cable
with common shield and individually shielded
cores
Conductor resistance
≤50 Ω/km (0.015 Ω/ft)
Capacitance: core/shield
≤420 pF/m (128 pF/ft)
Cable length (max.)
20 m (65 ft)
Cable lengths (available for
order)
5 m (15 ft), 10 m (32 ft), 20 m (65 ft)
Operating temperature
Depends on the device version and how the cable is installed:
• Standard version:
– Cable - fixed installation: –40 to +105 °C (–40 to +221 °F)
– Cable - movable: –25 to +105 °C (–13 to +221 °F)
• Order code for "Test, certificate", option
JP
:
– Cable - fixed installation: –50 to +105 °C (–58 to +221 °F)
– Cable - movable: –25 to +105 °C (–13 to +221 °F)
• Order code for "Test, certificate", option
JQ
:
– Cable - fixed installation: –60 to +105 °C (–76 to +221 °F)
– Cable - movable: –25 to +105 °C (–13 to +221 °F)
1)
UV radiation can impair the cable outer sheath. Protect the cable from direct sunshine where possible.
7.1.3
Terminal assignment
Transmitter: supply voltage, input/outputs
The terminal assignment of the inputs and outputs depends on the individual order
version of the device. The device-specific terminal assignment is documented on an
adhesive label in the terminal cover.
Supply voltage
Input/output
1
Input/output
2
Input/output
3
Input/output
4
1 (+)
2 (–)
EtherNet/IP
(RJ45 Stecker)
24 (+)
25 (–)
22 (+)
23 (–)
20 (+)
21 (–)
Device-specific terminal assignment: adhesive label in terminal
cover.
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected
by a connecting cable. The cable is connected via the sensor connection housing and the
transmitter housing.
Terminal assignment and connection of the connecting cable:
• Proline 500 – digital→ 45
• Proline 500 → 56
7.1.4
Device plugs available
Device plugs may not be used in hazardous areas!