Commissioning
Proline Promass F 500 EtherNet/IP
188
Parameter overview with brief description
Parameter
Description
User interface / User entry
Factory setting
Operating time
Indicates how long the device has been in
operation.
Days (d), hours (h), minutes
(m) and seconds (s)
–
Reset access code
Reset access code to factory settings.
For a reset code, contact your
Hauser service organization.
The reset code can only be entered via:
• Web browser
• DeviceCare, FieldCare (via service
interface CDI-RJ45)
• Fieldbus
Character string comprising
numbers, letters and special
characters
0x00
Using the parameter to reset the device
Navigation
"Setup" menu → Advanced setup → Administration
Parameter overview with brief description
Parameter
Description
Selection
Factory setting
Device reset
Reset the device configuration - either
entirely or in part - to a defined state.
• Cancel
• To delivery settings
• Restart device
• Restore S-DAT backup
Cancel
10.7 Simulation
The
Simulation
submenu enables you to simulate, without a real flow situation, various
process variables in the process and the device alarm mode and to verify downstream
signal chains (switching valves or closed-control loops).
Navigation
"Diagnostics" menu → Simulation
‣
Simulation
Assign simulation process variable
Process variable value
Status input simulation
Input signal level
Current input 1 to n simulation