2. General
2-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2004-10
2.3.6 Temperature
Control
Peltier-based temperature control.
PARAMETERS CHARACTERISTICS
Temperature range:
Room temperature up to 42°C increments of 0.1°C
Accuracy
+/- 0.2°C
Preheating time
20 min.
The
sunrise
specification of the temperature control unit (ac/- 0.2°C)
applies if everything (reader, microplate, reagents, samples) has already target
temperature.
There is no specification for the heating up period and we are aware that within
this period not all wells are heated up equally. This effect will be seen especially
when reagents are not preheated.
During preheating the plate transport with a 96 well microplate loaded has to be
inside of the instrument. Only in this way a homogenous temperature is received
in the instrument. The microplate must not be the same as the one used for
measurement but should be clean to keep influences on the following
measurement low.
Every time the plate transport is moved it will take up to 1 min to reach again a
homogenous temperature inside of the instrument.
Bear in mind that if non-preheated microplate and/or reagents are added
temperature control will take a certain time depending on the temperature
difference between microplates.
2.3.7
Barcode Laser Scanner
WARNING
LASER RADIATION – DO NOT STARE INTO THE BEAM!
CLASS II / 2 LASER PRODUCT.
The Class II / 2 Laser Scanner corresponds to the following norms:
•
DIN EN 60825-1 : 2001
•
CDRH 21 CFR 1040.10
PARAMETERS CHARACTERISTICS
Classification
Class II / 2 Laser Product
Input power
5 V DC +/- 10%
Emission duration
> 0.25 s
Scan rate
42 +/- 3 Scans / second (bi-directional)
Laser power
< 1 mW
Definition of depth
Max. 40 cm
Resolution 0.15
cm
Min. print contrast
25% at 675 nm
Ambient light
Sun light: 40000 Lux
Halogen light: 1500 Lux