7. Mechanical System
2004-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
7-5
7.6 Transport Box Adjustments
After the transport system and all other parts have been replaced, to ensure
correct alignment of the transport system the following adjustments should be
performed using the Sunset Software (as described in chapter 4.4 SunSet
Software (Service tool kit is necessary!)).
•
Plate Calibration
•
Measurement Offset
•
Duration Test
7.7 Replacing the Temperature Control Unit
1. Switch the instrument OFF
2. Remove the main power cable from the socket
3. Remove top cover and upper EPAC (see 5 Removing and Replacing the
Instrument Cover)
4. Remove front cover (see 5 Removing and Replacing the Instrument Cover)
5. Disconnect ST1, ST8, ST7, ST3, ST15, ST20 and ST 21
from Mainboard and all connections to the Powerboard
6. Remove the Power Board and Mainboard
7. Disconnect the Fiber bundle from the Filter Guide
8. Disconnect the Thermo Board and remove the Transport Box carefully
9. Remove the foam rubber cover from the backside of the instrument
10. Disconnect the Flex cable from the Thermo Board on the left side and
remove the nozzle plate.