2. General
2004-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2-7
2.3.2
Standard Instrument (Standard Filter Carriage)
PARAMETERS CHARACTERISTICS
Measurement time
:
dual wavelength
single wavelength
8 seconds
6 seconds
Wavelength range
Standard
340 - 750 nm
Filter wavelength accuracy
Central wave/- 2 nm
Filter bandwidth
At 50% transmission:
10 +/- 2nm
Measurement range
:
340 – 399 nm
400 – 750 nm
0 - 3.000 Abs.
0 - 4.000 Abs.
Resolution
: 0.001
Abs.
Accuracy
:
(492 nm) 0.000 - 2.000 Abs.
(492 nm) 2.000 - 3.000 Abs.
Better than +/- 1.0 % +/- 0.010 Abs.
Better than +/- 1.5 % +/- 0.010 Abs.
Precision
:
(492 nm) 0.000 - 2.000 Abs.
(492 nm) 2.000 - 3.000 Abs.
Better than +/- 0.5 % +/- 0.005 Abs.
Better than +/- 1.0 % +/- 0.005 Abs.
Linearity
:
(340 - 399 nm) 0.000 - 2.000 Abs.
(400 - 750 nm) 0.000 - 2.000 Abs.
2.000 - 3.000 Abs.
Better than +/- 2 %
Better than +/- 1 %
Better than +/- 1.5 %
Wavelength selection
:
Standard filter
Narrow band interference filters. Up to
four filters can be mounted in a filter
carriage. The instrument can use up to
eight different filter carriages.
Light source:
Halogen lamp 20 W
All connected devices must be approved and listed as per EN 60950, UL 1950
or CSA C22.2 No. 950 for Data Processing Devices.
Computer interface:
Serial RS 232 C
Printer
300 - 38,400 baud
Only with the Touchscreen option.