11. Maintenance & Cleaning
11-8
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2004-10
11.6 Preventive Maintenance Plan for
sunrise
This preventive maintenance plan is for standard throughput instruments. For
instruments that are used in high throughput, the maintenance intervals might be
shorter.
11.6.1 Daily
•
No daily maintenance is required.
11.6.2 Weekly
•
Clean the cover and the plate transport with a mild detergent.
Caution
Never use Acetone as it will damage the covers.
11.6.3 Every
Six
Months
•
Clean the filters using an optical cleaning solution (lens tissue
recommended)
11.6.4 Yearly (Customer or Service Technician)
Either done by customer:
•
Perform the QC Pac 2 Test (see QC Pac Instructions for Use)
Or done by service technician:
•
Perform extended operational quality check
11.6.5 Every Four Years (Service Technician Required)
•
Replace the lamp and the filters
•