2. General
2004-10
Technical Manual for sunrise RC, TW, TS, TC, BC, 6F No. T 137 302 Rev No. 1.1
2-1
2. General
2.1 Introduction
Note
The
sunrise
Absorbance Reader is intended for use with only external
software.
Caution
Before starting measurements, make sure that the microplate
position A1 is inserted correctly.
The
sunrise
instruments are fully automatic, microprocessor controlled readers
designed for professional use for in vitro diagnostic applications, enabling the
user to measure the light absorbance (optical density) of samples in 96 well
microplates according to the specifications described in this publication.
Note
Results obtained using the
sunrise
are influenced by the proper use of the
instrument and microplates, according to the instructions given in this
document, as well as the liquid compounds used (reagents, chemistry). The
instructions for use, for storage and for other manipulations in connection
with samples or reagents must be strictly followed. Taking this fact into
consideration, results must be interpreted carefully.
By reading twelve wells simultaneously, the instrument is able to measure a
microplate in approximately six seconds.
The obtained transmission values are converted into OD values according to the
following formula:
T
1
Log
OD
I
I
T
on
Transmissi
o
=
=
light
detected
I
light
emitted
I
o
=
=
With an innovative range of options, this versatile microplate reader gives
diagnostic and research laboratories all the features for numerous purposes.
Based on a new design concept EPAC, the excellent optical performance and
high quality of the
sunrise
will guarantee fast, reproducible and accurate
measurements.
The
sunrise
is designed to be fitted into Tecan robotic systems.
The
sunrise
Absorbance Reader is an optical 96-well microplate reader for the
measurement of absorbance and turbidimetry of samples from the biological or
non-biological origin according to the specifications described in this document,
intended for professional use only.