5. Operating the Instrument
98
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
5.6.5
Measurement with a Simultaneous G-Factor Calibration
When
Calibrate
is selected, the G-factor is determined for the current
measurement parameters and used for the following FP measurement. In order
to perform the G-Factor calibration, please define:
Reference value
: select a polarization value for the reference used, e.g. 20 mP
for a 1 nM Fluorescein solution in 0.01 M NaOH.
Reference range
: click
Change
and select the wells filled with the reference.
Blank range
: click
Change
and select the wells filled with the reference blank.
Select
Same as measurement blank
if the reference blank is the same as the
measurement blank.
Note
By filling in more than one well with polarization references and
reference blanks, the mean values will be calculated and therefore the
calibration result will be more accurate.