6. Instrument Features
146
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
2016-04
6.11 Cuvette Features (Infinite M200 PRO
only)
Note
The option ‘Cuvette’ is only available for the Infinite M200 PRO. This
module cannot be installed in an Infinite F200 PRO.
The cuvette option allows for performing absorbance measurement in fixed
wavelength and scan mode.
Parameters
Characteristics
Wavelength Range
230 – 1000 nm
(no filter necessary,
selection in 1 nm steps possible)
Measurement Range
0 – 4 OD
6.11.1 Cuvette Specifications
The following specifications are valid for the wavelength range from 300-700 nm,
number of flashes 25:
Cuvette types
12.5 x 12.5 mm:
maximum height 55 mm (including lid);
micro cuvettes:
center height 15 mm,
measurement window > 2 mm (diameter).
Accuracy 0 – 2 OD
<
±
(1 % + 18 mOD)
Accuracy 2 – 3 OD
<
±
2.5 %
Reproducibility 0 – 2 OD
<
±
(1 % + 10 mOD)
Reproducibility 2 – 3 OD
<
±
2.5 %
Linearity 0 – 2 OD
R
2
> 0.998
Baseline Flatness
±
10 mOD (1 sigma)
CAUTION
THE CUVETTE PORT OF INFINITE M200 PRO CANNOT BE USED FOR
CUVETTES WITH A MEASUREMENT WINDOW < 2 MM (DIAMETER)
AND A CENTER HEIGHT BELOW 15 MM.
STOP