9. Troubleshooting
2016-04
IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6
185
9. Troubleshooting
Error #
Error Text
Description
1
Command is not valid
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
2
Parameter out of range
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
3
Wrong number of parameters
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
4
Invalid parameter
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
5
Invalid Parameter at pos
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
6
[prefix] is missing
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
7
RS485 Timeout at module
[module descr]
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
8
Invalid module number [Nr]
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
9
Binary Transfer command:
[cmd] at module [n]
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
10
Error at command [cmd] at
module [n],
Unspecific internal communication error. Please report this
error to your local Tecan customer support office.
11
LID open
Plate transport or filter slide lid were open during a
measurement or the instrument was used in very bright
environment (<< 500 LUX). Please check if the lid closes
completely or if the environment was too bright.
12
LUMI FIBER broken
Hardware Failure Luminescence Module. Please report this
error to your local Tecan customer support office.
13
Z Motor out of Safety-Range
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
14
Filter is not defined
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
15
X drive init error
Hardware Failure Plate Transport Module. Please report
this error to your local Tecan customer support office.
16
Y drive init error
Hardware Failure Plate Transport Module. Please report
this error to your local Tecan customer support office.
17
z drive init error
Hardware Failure z-drive Module. Please report this error to
your local Tecan customer support office.