Page 18
8.
Press the ENTER key to start the auto range measurement.
9.
To decide the auto range setting with the measurement result
Press the STOP key after taking a necessary time for auto range measurement. The system
automatically sets the suitable input range to each channel based on the measurement.
To cancel the auto range measurement
Press the left key. The system cancels the auto range measurement and does not set the input
range to each channel.
10.
To decide the auto range setting with the measurement result
Press the ENTER key to display the selected input range of each channel. Press the up and down
keys to locate the channels to display. Press the ENTER key after confirming the selections.
<INPUT AMP> is displayed.
To cancel the auto range measurement
Press the ENTER key to return to <AUTO RANGE> without applying the measured auto range
setting.
About Zero Shift at Auto Range
Adjust the zero shift again after operating the auto range. The auto range measures the signal
by canceling the preset zero shift value.
If the auto range results to change the input range, adjust the zero shift percentage according to
the specified input range since the zero shift specifies its value in percentages against the input
range.
Caution on Using External Sampling Clock at Auto Range
When you set the sampling to external, apply an external sampling clock before starting the auto
rage function.
Содержание DS Series
Страница 7: ...Page v Note...
Страница 8: ...Page vi Note...
Страница 14: ...Page 6 Note...
Страница 22: ...Page 14 Note...
Страница 30: ...Page 22 6 Press the ENTER key to return to INPUT AMP...
Страница 76: ...Page 68 Note...
Страница 78: ...Page 70 Note...
Страница 86: ...Page 78 Note...
Страница 98: ...Page 90 Note...
Страница 100: ...Page 92 Note...
Страница 102: ...Page 94 Note...
Страница 103: ...DS Series Instruction Manual 2006 12 Page 95 Note...