TimeAnalyzer 7x00 User’s Guide
098-00208-000 Revision A – February, 2010
Chapter 3 Configuring
Configuring TP5000 Module Using TimeMonitor PDV
6. Type in the desire access-level (user, config, or admin) and press
Enter
. The
system prompts you for a password.
7. Type a password and press
Enter
. The password you enter will be the log-in
password for the user. The system responds with the confirm password prompt.
8. Retype the password and press
Enter
. The system indicates that “
CLI user has
been added: <user>
”.
9. Type
show user
and press
Enter
to verify that the account exists.
10.Repeat Steps 3 through 9 to modify other user passwords and access levels.
Saving and Restoring Configuration Data
Symmetricom recommends that you keep an electronic version and/or a written
version of the provisioning changes you make to the TP5000 probe module.
The configuration is automatically backed up one minute after a configuration
change.
To backup up the current configuration immediately, enter the following:
1. Type
set configuration backup local
and press
Enter
.
Example
To change the current configuration to factory default while preserving user and IP
information, enter the following:
1. Type
set configuration default
and press
Enter
.
Example
To change the current configuration to the factory default, enter the following:
1. Type
set configuration factory
and press
Enter
.
Notes:
Setting the configuration to
default
will not overwrite the user login
information or the system’s IP address. Setting the configuration to
factory
will overwrite the user login information and return the
system’s IP address to the factory set address.