TimeAnalyzer 7x00 User’s Guide
43
Chapter 2 Setting Up
Normal Power Up Indications
IOC Alarm
IOC module alarm/fault indicator
ALM
Off
- IOC operating normally
Amber
- IOC operational Minor Alarm(s)
Red
- IOC has Failed or has Major/Critical Alarm(s)
Flashing Green
- Software starting up
Ethernet Port 1
ETH1 port that is used for timing
distribution
ETH1
Green
- Link has been established
Flashing Green
- Link activity
Off
- Link has not been established
Ethernet Port 2
ETH2 port that is used for timing
distribution
ETH2
Green
- Link has been established
Flashing Green
- Link activity
Off
- Link has not been established
Local Oscillator State
Indicates the current mode of
operation of the LO servo
HOLD
Off
- LO servo is operating in Locked mode
Amber
- LO servo is operating in Holdover or
Acquire modes
Flashing Amber
- LO servo is operating in
Warm-up mode
Note: This LED will remain Amber when using the
TP5000 probe module in free-run mode. To avoid a
Red LED for GPS and a resulting alarm, set the
Time of Day source to “sys”.