AMI Deltacon Power
Troubleshooting
49
Error
Description
Corrective action
E001
Cond. 1 Alarm high
– check process
E002
Cond. 1 Alarm low
– check process
E003
Cond. 2 Alarm high
– check process
– check programmed value, see
E004
Cond. 2 Alarm low
– check process
– check programmed value, see
E007
Temp. 1 high
– check process
– check programmed value, see
E008
Temp. 1 low
– check process
– check programmed value, see
E009
Sample Flow high
– check sample inlet pressure
– check programmed value, see
E010
Sample Flow low
– check sample inlet pressure
– Check flow regulating valve
– check programmed value, see
E011
Temp. 1 shorted
– Check wiring of temperature sensor
– Check temperature sensor
E012
Temp. 1 disconnected
– Check wiring of temperature sensor
– Check temperature sensor
E013
Case Temp. high
– check case/environment temperature
– check programmed value, see
E014
Case Temp. low
– check case/environment temperature
– check programmed value, see