A-96.250.451 / 190121
AMI Deltacon Power
Program List and Explanations
61
4.3 Logger
The instrument is equipped with an internal logger. The logger data
can be copied to a PC with an USB stick if option USB interface is in
-
stalled.
The logger can save approx. 1500 data records. Records consist of:
Date, time, alarms, measured value, measured value uncompensat
-
ed, temperature, flow.
Range: 1 Second to 1 hour
4.3.1
Log Interval:
Select a convenient log interval. Consult the table be
-
low to estimate the max logging time. When the logging buffer is full,
the oldest data record is erased to make room for the newest one
(circular buffer).
4.3.2
Clear Logger:
If confirmed with
yes
, the complete logger data is de
-
leted. A new data series is started.
4.3.3
Eject USB Stick
: With this function all logger data are copied to the
USB stick before the USB stick is deactivated.
Only visible it the optional USB interface is installed.
4.4 Display
Process values are displayed on two screens. Toggle screens with
the [
] key. Each screen displays max. 3 process values.
4.4.1
Screen 1
4.4.1.1
Row 1
4.4.1.2
Row 2
4.4.1.3
Row 3
Possible settings for all rows are:
None
Cond 1 (sc)
Cond 2 (cc)
Difference
pH (if <Calculations> = yes)
Ammonia (depends on the settings in <Sensor parameters>/
<Temp. comp.>)
Interval
1 s
5 s
1 min
5 min
10 min
30 min
1 h
Time
25 min
2 h
25 h
5 d
10 d
31 d
62 d