AMI Deltacon Power
Program List and Explanations
71
5.3.1.3
Sample Flow:
Define at which sample flow an alarm should be is
-
sued.
5.3.1.3.1
Flow Alarm:
Program if the alarm relay should be activated if there is
a flow alarm. Choose between yes or no. The flow alarm will always
be indicated in the display, pending error list, saved in the message
list and the logger. Available values: Yes or no
Note:
Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.3.2
Alarm High:
If the measuring values rises above the programmed
value E009 will be issued.
Range: 9–20 l/h
5.3.1.3.35
Alarm Low:
If the measuring values falls below the programmed val
-
ue E010 will be issued.
Range: 5–8 l/h
5.3.1.4
Case Temp.
5.3.1.4.1
Alarm high:
Set the alarm high value for temperature of electronics
housing. If the value rises above the programmed value E013 is is
-
sued.
Range: 30–75 °C
5.3.1.4.2
Alarm low:
Set the alarm low value for temperature of electronics
housing. If the value falls below the programmed value E014 is is
-
sued.
Range: -10 to +20 °C
5.3.2 and 5.3.3
Relay 1 and 2:
The contacts can be set as normally open or normal
-
ly closed with a jumper. See
.
The function of relay contacts 1 or 2 is defined by the user.
Note:
The navigation in the menu <Relay 1> and <Relay 2> is
equal. For reason of simplicity only the menu numbers of Relay 1
are used in the following.
1
First select the functions as:
– Limit upper/lower
– Control upwards/downwards
– Timer
– Fieldbus
2
Enter the necessary data depending on the selected function.
The same values may also be entered in menu