A-96.250.451 / 190121
AMI Deltacon Power
Maintenance
40
7
Navigate to menu <Maintenance>, <Change of Resin>, press
[Enter] and set it to <Yes>.
8
Open and adjust the sample flow.
9
Pre-rinse the new cation exchanger resin until the display shows
stable measuring values.
with pre-rinse
option
Step 1 to 3 same procedure as in
“without pre-rinse option”
:
4
Unscrew and carefully remove the pre-rinsed cation exchanger
bottle [D] from the pre-rinse holder [C].
5
Carefully, without spilling water, push the cation exchanger bottle
[D] over the inlet filter holder [E] into the bottle holder [B].
6
Screw the cation exchanger bottle into the bottle holder.
Do not tighten the bottle too firmly, this could damage the
gasket.
7
Install a new bottle with fresh, unused resin into the pre-rinse bot
-
tle holder [C].
The new c
.
ation exchanger resin will be pre-rinsed and ready
for use if the next exchange is necessary.
A
B
C
D
E
F
Flow cell
Bottle holder
Pre-rinse bottle holder
Pre-rinsed cation
exchanger bottle
Inlet filter holder
Exhausted cation
exchanger bottle
A
B
E
C
D
F