A-96.250.451 / 190121
AMI Deltacon Power
Product Description
14
2.2. Instrument Specification
Power Supply
AC variant:
100–240 VAC (±10%)
50/60 Hz (±5%)
DC variant:
10–36 VDC
Power consumption:
max. 35 VA
Sample
requirements
Flow rate:
5–20 l/h
Temperature:
up to 50 °C
Inlet pressure:
up to 2 bar
Outlet pressure:
pressure free
On-site
requirements
The analyzer site must permit connections to:
Sample inlet:
Swagelok 1/4” adapter for stainless
steel tube
Sample outlet:
G ½” adapter for flexible tube
15 x 20 mm
Measuring
range
Measuring range
Resolution
0.055 to 0.999
S/cm
0.001
S/cm
1.00 to 9.99
S/cm
0.01
S/cm
10.0 to 99.9
S/cm
0.1
S/cm
100 to 1000
S/cm
1
S/cm
1.00 to 2.99 mS/cm
0.01 mS/cm
3.0 to 9.9 mS/cm
0.1 mS/cm
10 to 30 mS/cm
1 mS/cm
Automatic range switching.
Accuracy
±1% of measuring value (up to 5 mS/cm)
±3% of measuring value (up to 30 mS/cm)
Transmitter
specifications
Housing:
Aluminium with a protection degree of
IP 66 / NEMA 4X
Ambient temperature:
-10 to +50 °C
Storage and transport:
-30 to +85 °C
Humidity:
10–90% rel., non-condensing
Display:
backlit LCD, 75 x 45 mm