A-96.250.451 / 190121
AMI Deltacon Power
Program List and Explanations
69
5.3.1.1.1.45
Delay:
Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.1.2
Cond. 2 (cc)
5.3.1.1.2.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E003, is displayed in the message
list.
Range: 0.000 –3000
S
5.3.1.1.2.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.000 –3000
S
5.3.1.1.2.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value.
Range. 0.000 –3000
S
5.3.1.1.2.45
Delay:
Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
5.3.1.1.4
pH
(if Calculations = yes)
5.3.1.1.4.1
Alarm High:
If the measured value rises above the alarm high value,
the alarm relay is activated and E033, is displayed in the message
list.
Range: 0.00–14 pH
5.3.1.1.4.25
Alarm Low:
If the measured value falls below the alarm low value,
the alarm relay is activated and E034 is displayed in the message
list.
Range: 0.00–14 pH
5.3.1.1.4.35
Hysteresis:
Within the hyst. range, the relay does not switch. This
prevents damage of relays contacts when the measured value fluctu
-
ates around the alarm value.
Range: 0.00–14 pH
5.3.1.1.4.45
Delay:
Duration, the activation of the alarm relay is retarded after the
measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec