A-96.250.671 / 070222
AMI Silica
Program List and Explanations
79
2.4 I/O State
Shows actual status of all in- and outputs.
2.4.1 and 2.4.2
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
3 Maintenance
3.1 Calibration
3.1.5
Performs a calibration using the standard solution. Follow the in
-
struction on the screen. See
3.2 Service
3.2.1
Verification:
Performs a verification using the verification kit. Follow
the instruction on the screen. See
3.2.2
Fill System:
Activates the reagent pump and fills all tubes from the
container to the 6-way valve.
3.2.3
Manual Filling:
In the menu <Manual filling> the position of the
6
-
way valve can be set manually and the peristaltic pump can be
switched on and off.
3.2.3.1
Position
: Set the position of the 6
-
way valve.
3.2.3.2
Pump
: Switch the peristaltic pump on or off.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Input:
Open or closed
Signal Output 1 and 2:
Actual current in mA
Signal Output 3:
Actual current in mA (if option is installed)
Pos. 1 Reagent 1
Pos. 2 Reagent 2
Pos. 3 Reagent 3
Pos. 4 Reagent 4
Pos. 5 Loop
Pos. 6 –