A-96.250.671 / 070222
AMI Silica
Maintenance
53
6.5. Calibration
Preparing
the standard
Swan offers a 100 ppm stock solution, from which you can produce
your own standard. By default, the instrument is programmed for a
standard of 100 ppb. Other concentrations can be programmed in
menu <Installation>/<Sensors>/<Standard>.
To prepare a 100 ppb standard solution, dilute 1 ml of stock solution
with 1 l of demineralized water.
Calibration
1
Navigate to <Maintenance>/<Calibration>.
2
Follow the instructions on the screen.
Note:
If an AMI Sample Sequencer is installed, the flow
regulating valve [B] on the flow cell block of the constant head
must be closed during the calibration. Otherwise, the standard
solution may flow back into the sample feed line.
A
B
Flow cell block of the
constant head
Flow regulating valve
Level
A
B