A-96.250.671 / 070222
AMI Silica
Program List and Explanations
77
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active, acknowl
-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5
Demands necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1
Shows the error history: Error code, date / time of issue and status
(active, acknowledged, cleared). 65 errors are memorized. Then the
oldest error is cleared to save the newest error (circular buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
Designation of the instrument.
Version
: Firmware of the instrument (e.g. V6.22-01/22)
2.1.3
Peripherals
:
2.1.3.1
PeriClip
: Firmware of peristaltic pump (e.g. 1.06)
RoValve
: Firmware of rotary valve (6-way valve) (e.g. 1.60)
2.1.4
Factory Test
: Test date of instrument and motherboard.
2.1.5
Operating Time:
Years / days / hours / minutes / seconds.
2.2 Sensors
2.2.1
FOME Sensor:
Current value:
Shows the actual photometer signal in ppb.
Raw value:
Shows the actual photometer signal in V.
Absorbance
: Process value, depends on sample.
2.2.1.4
Cal. History:
Review diagnostic values of the last calibrations.
Factor (slope).
Max. 64 data records are memorized. One calibration step corre
-
sponds to one data record:
Slope photometer: 0.5–2.0