A-96.250.671 / 070222
AMI Silica
Maintenance
60
Dismount
pump tubes
The pump tube can easily be dismounted and mounted. Proceed as
follows:
1
Switch off the instrument according to instructions in
eration for Maintenance, p. 48
2
Remove the protection cap.
3
Open the occlusion frame [B] by turning it counter-clockwise.
4
Remove the pump tube [D] from the rotor [C] by pulling the com
-
plete occlusion frame [B] out of the holder.
5
Disconnect the reagent tubes from the old pump tube and con
-
nect them to the new pump tube
6
Install the new pump tube by pushing the occlusion frame onto
the holder.
7
Lock the occlusion frame. Check that the occlusion frame and the
tube are aligned perpendicular to the axis of the rotor.
8
Insert the suction lances into the corresponding containers.
9
Start the <Fill system> function.
A
B
C
D
E
F
Pump housing
Occlusion frame open
Rotor
Pump tube
Pump inlet
Pump outlet
A
B
D
C
E
F