A-96.250.671 / 070222
AMI Silica
Troubleshooting
67
Error
Description
Corrective action
E001
Si 1 Alarm high
– check process
– check programmed value
E002
Si 1 Alarm low
– check process
E003
Si 2 Alarm high
– check process
– check programmed value
E004
Si 2 Alarm low
– check process
E007
Sample Temp. high
– check process
– check programmed value
E008
Sample Temp. low
– check process
– check programmed value
E009
Sample Flow high
– check inlet pressure
– re-adjust sample flow
– check programmed value
E010
Sample Flow low
– check inlet pressure
– re-adjust sample flow
– clean instrument
E012
Temp. disconnected
– shut off power
– check wiring of photometer, see
E013
Case Temp. high
– check case/environment temperature
– check programmed value
E014
Case Temp. low
– check case/environment temperature
– check programmed value
E015
Valve defective