A-96.250.671 / 070222
AMI Silica
Instrument Setup
40
Instrument
with AMI Sam
-
ple Sequencer
If an AMI Sample Sequencer is installed, proceed as follows:
1
Switch on power.
2
Open the flow regulating valve [D] on the flow cell block of the
constant head.
3
Adjust the sample flow using the flow regulating valves [F] of the
AMI Sample Sequencer.
4
Fill or Flush Reagent System, p. 62
5
Check tube connections and flow cell for leaks and repair if
necessary.
6
Let the instrument run-in for one hour.