A-96.250.671 / 070222
AMI Silica
Instrument Setup
39
Two-channel
instrument
If the second sample stream option is installed, proceed as follows:
1
Switch on power.
2
Adjust the sample flow using the flow regulating valves [D] of the
second sample stream option.
3
Start <Fill system>, see
Fill or Flush Reagent System, p. 62
.
4
Check tube connections and flow cell for leaks and repair if
necessary.
5
Let the instrument run-in for one hour.
A
B
C
D
E
Cover
Outer tube
Blind plug
Flow regulating valves
Second sample stream
option
Level
E
D
2
1
C
A
B