Troubleshooting
66
7.3. Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is
exceeded.
Such Errors are marked
E0xx
(bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are
recovered (i.e. Sample Flow low).
Such Errors are marked
E0xx
(
bold and orange
)
Errors which indicate a hardware failure of the instrument.
Such Errors are marked
E0xx
(
bold and red
)
Error or
fatal Error
Error not yet acknowledged.
Check
Pending Errors 1.1.5
and take
corrective action.
Reagent level low
Indicates the remaining reagent in
percent
Navigate to menu <Messages>/
<Pending Errors>.
Press [ENTER] to acknowledge the
Pending Errors.
The Error is reset and saved in
the Message List.
23 B/s
HOLD
14:10:45
R1
21.5 ppb
R2
1.1
Maintenance List
Pending Errors
Messages
Message List
1.1.5
Pending Errors
Error Code
E002
Si1 Alarm low
<Enter> to Acknowledge