A-96.250.671 / 070222
AMI Silica
Installation
24
3
Unscrew and remove the flow regulating valve [E] from the flow
cell block [F].
4
Screw the blind plug [D] into the flow cell block.
5
Remove the sample inlet tube [N] from the sample inlet (elbow
union) [G].
6
Remove the elbow union from the flow cell block.
7
Put one of the O-rings [C] into the bore of the block housing [A]
and the other one into the bore of the flow cell block [F].
8
Insert the connecting piece [B] into the bore of the flow cell block.
9
Slide the block housing [A] over the connecting piece and press it
against the flow cell block [F] while screwing the block housing
with the 2 fixing screws [L] onto the panel.
10
Push a 15 x 20 mm tube [K] (not included in the installation set)
over the ½” hose nozzle of the block housing and put the end of
the tube into a pressure free drain.
11
Connect the sample stream 1 [I] and the sample stream 2 [J] to
the corresponding sample inlets of the block housing according
to section
.