![Swan Analytical Instruments AMI Codes-II O3 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-codes-ii-o3/ami-codes-ii-o3_operators-manual_1411369074.webp)
A-96.250.881 / 070622
AMI Codes-II O3
Program List and Explanations
74
As process
values
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
5.2.1.40
Scaling:
Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter DIS:
5.2.1.40.10
Range low
: 0.0–500.0 ppb
5.2.1.40.20
Range high:
0.0–500.0 ppb
Parameter Sample flow
5.2.1.40.13
Range low
: 0 –600 B/s
5.2.1.40.23
Range high
: 0 –600 B/s
A
B
linear
bilinear
X
Measured value
X
Measured value (logarithmic)
20
0.0
0.1
0.2
0.3
0.4
0.5
10 12
(0 - 20 [mA])
0 / 4
(4 - 20 [mA])
[mA]
X
A
B
20
1
0
1
2
3
4
10
100
1’000
10’000
10 12
(0 - 20 [mA])
0 / 4
4
2
6
(4 - 20 [mA])
[mA]
X