AMI Codes-II O3
Troubleshooting
60
Error
Description
Corrective action
E001
DIS. Alarm high
– check process
– check programmed value in menu
E002
DIS. Alarm Low
– check process
– check programmed value in menu
E005
DIS. too high
– check process
E009
Sample Flow high
– check sample input pressure
– readjust sample flow
– check programmed value in menu
E010
Sample Flow low
– check sample input pressure
– readjust sample flow
– check programmed value in menu
E013
Case Temp. high
– check case/environment temperature
– check programmed value in menu
E014
Case Temp. low
– check case/environment temperature
– check programmed value in menu
E017
Control Timeout
– check control device or programming in
Installation, Relay contact, Relay 1/2
E018
Reagent Pump
– shut off power
E019
Photometer not
connected
– shut off power
– check wiring of photometer, see
E020
Photometer dirty
– check process,
– clean photometer, see