![Swan Analytical Instruments AMI Codes-II O3 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-codes-ii-o3/ami-codes-ii-o3_operators-manual_1411369055.webp)
A-96.250.881 / 070622
AMI Codes-II O3
Maintenance
55
Dismount
pump tubes
The pump tube can easily be dismounted and mounted. Proceed as
follows:
1
Switch off the instrument according to instructions in
eration for Maintenance, p. 42
.
2
Remove the protection cap.
3
Open the occlusion frames [B] by turning them counter-clock
-
wise.
4
Remove the pump tubes [D] from the rotor [C] by pulling the com
-
plete occlusion frames [B] out of the holder.
5
Disconnect the reagent tubes from the old pump tubes and con
-
nect them to the new pump tubes.
6
Install the new pump tubes by pushing the occlusion frames onto
the holder.
7
Lock the occlusion frames. Check that the occlusion frames and
the tubes are aligned perpendicular to the axis of the rotor.
8
Insert the suction lances into the corresponding containers.
9
Start the <Fill system> function.
A
B
C
D
E
F
Pump housing
Occlusion frame open
Rotor
Pump tube
Pump inlet
Pump outlet
A
B
D
C
E
F