A-96.250.881 / 070622
AMI Codes-II O3
Maintenance
53
Cleaning
1
Switch off the instrument according to instructions in
eration for Maintenance, p. 42
2
Remove the constant head cover [A].
3
Remove the constant head tube [C] from the flow cell block.
4
Pull the overflow tube [B] out of the flow cell block [D].
5
Clean all acrylic parts with a soft brush (bottle cleaner) and soapy
water.
6
Rinse well with demineralized water.
6.7.2
Assemble the Flow Cell
1
Replace the gasket [D] before reassembling the flow cell.
Note:
A film of teflon paste (e.g. Fomblin from Solvay Solexis)
on the gaskets improves tightness and life time.
2
Push the overflow tube [B] through the flow cell block as far as it
reaches the drain.
3
Install the constant head tube [C] onto the flow cell block.
4
Put the cover onto the constant head tube.
5
Align the overflow tube with the upper level mark.
A
B
C
D
E
Constant head cover
Overflow tube
Constant head tube
Gasket
Flow cell block
Level
A
B
E
C
D