A-96.250.881 / 070622
AMI Codes-II O3
Program List and Explanations
70
2.2.1.4
Cal. History:
Shows the diagnostic values of the last calibrations.
Number
: Calibration counter.
Date, Time
: Date and time of the calibration.
Slope
: Slope is a correction factor calculated on the basis of a
process calibration.
Range: 0.8–1.2
2.2.1.5
Ver. History:
Shows the verification values of the last verifications:
Number
: Verification counter.
Date, Time
: Date and time of the verification.
Absorbance
: Measured absorbance of the reference kit.
Reference value
: True value of the reference kit according to label.
Deviation:
Deviation between the measured value and the reference
value in %.
2.2.3
Miscellaneous
:
2.2.3.1
Case Temp:
Shows the current temperature in [°C] inside the
transmitter.
2.3 Sample
2.3.1
Sample ID:
Shows the identification assigned to a sample. This
identification is defined by the user to identify the location of the
sample.
Sample Flow:
Shows the actual sample flow in B/s (bubbles per
second)]. The Sample flow must be above 5 B/s.
2.4 I/O State
Shows current status of all in- and outputs.
2.4.1/2.4.2
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
Alarm Relay:
Active or inactive
Relay 1 and 2:
Active or inactive
Input:
Open or closed
Signal Output 1 and 2:
Actual current in mA
Signal Output 3:
Actual current in mA (if option is installed)