AMI Codes-II O3
Product Description
14
Time
interval of a
measurement
The measuring interval can be set between 5 and 10 minutes. The
time sequence of a measurement with a measuring interval of 5 min
is shown in the diagram below.
The blue bar represents the sample which flows continuously
through the photometer. A short time before the measurement starts,
a zero point measurement is performed. Then the peristaltic pump
starts and a small portion of the reagents [J] and [K] is added. Shortly
after, when the mixture is in the photometer, the sample is measured.
J
K
OXYCON ON-LINE DPD reagent
OXYCON ON-LINE Ozone reagent
Zero point measurement
Sample measurement
1
0
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
min
J
K