A-96.250.881 / 070622
AMI Codes-II O3
Maintenance
44
Reagent
consumption
The 2 liter reagent canister will last for at least 33 days of operation
with the default measurement interval of 6 minutes. The provided re
-
agent set (for 3 canisters) therefore lasts for at least 3 months of op
-
eration.
Note:
To achieve a reagent blank of less than 5 ppb, the
canisters must be consumed within 30 days.
Contents
of the
reagent set
A-85.410.210 Reagent set for the measurement of ozone:
Oxycon On-Line DPD (3 bottles)
Oxycon On-Line Buffer O3 (3 bags)
Personal protective equipment:
Measuring interval
Duration per canister
5 minutes
28–40 days
6 minutes
33–47 days
8 minutes
44–63 days
10 minutes
55–79 days
Oxycon On-Line DPD:
H314: Causes severe skin burns and eye
damage.
H318: Causes serious eye damage.
Oxycon On-Line Buffer O3:
H373: May cause damage to organs through
prolonged or repeated exposure.