Instrument Setup
34
4.3. Establish Sample Flow
WARNING
Water pollution
The drain of the photometer outlet contains DPD.
At no means recirculate it into the water system.
1
Open the flow regulating valve [E] and wait until the flow cell is
completely filled.
2
Adjust overflow tube to the marked level.
3
Switch on power.
4
Adjust the sample flow so that always a small part of the sample
drains off through the overflow tube.
5
Fill or Flush Reagent System, p. 35