A-96.250.881 / 070622
AMI Codes-II O3
Maintenance
56
6.9. Longer Stop of Operation
1
Put the suction lances into a bucket with demineralized water.
2
Start <Fill system>.
The reagent tubes are flushed with water.
3
Remove the suction lance from the water.
4
Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5
Stop sample flow.
6
Wait until flow cell is empty.
7
Shut off power of the instrument.
8
Open the occlusion frames of the peristaltic pump, see
.