Maintenance
54
6.8. Tube Replacement
6.8.1
Replace the Pump Tubes
The pump tube [D] of the peristaltic pump is exposed to a minimal
wear. It is therefore recommended to exchange the pump tube annu
-
ally.
CAUTION
Pollution of reagents possible.
If the occlusion frames are opened during operation, already
mixed reagents will flow back into the reagent canisters and pol
-
lute the reagents.
Never open the occlusion frames if the instrument is in opera
-
tion.
Stop of Operation for Maintenance, p. 42
before opening the occlusion frames.
Overview
A
B
C
D
Pump housing
Occlusion frame closed
Rotor
Pump tube
E
F
G
Pump inlet
Pump outlet
Protection cap
A
B
D
C
E
F
G