General Use Procedure
30
When the Target Time Is Reached
Once the inflation timer has reached the target time set for the procedure,
the following happens:
1.
The
Alarm Indicator Mute
button flashes red, the ‘Time’
setting on the LCD blinks once per second and an audible
alarm chimes 6 times – over 6 seconds. The audible chime
then pauses 6 seconds and repeats the 6-beep sequence.
2.
The alarm chime will continue indefinitely until (a) the
Alarm
Indicator
Mute
button is pressed, (b) additional time is added
to the timer or (c) the cuff is deflated.
3.
When the
Alarm
Indicator
Mute
button is pressed, the
audible chime is silenced for 15 minutes. However, the time
value on the LCD display will continue to blink once per
second.
When the target time is reached, do one of the following:
♦
Deflate the cuff by pressing the
Deflate
button for 1.5 seconds,
and if necessary, re-inflate the cuff.
♦
Press the
Alarm Indicator Mute
button to mute the alarm for 15
minutes. The Alarm light will continue to flash red. The
SmartPump will continue to record and display the total elapsed
cuff time.
♦
Add more time using the
Set/Save
button and (
+
) button. See
Setting Timer Cuff Pressure.
N
OTE
:
The SmartPump does not automatically deflate when the target time is
reached. Deflation must be initiated by pressing the
Deflate
button.