User/Patient Safety
7
Precautions for Use
Extreme care must be taken when using tourniquets. Minimum cuff pressure
and application time should be used. Monitor time and pressure throughout the
procedure as injury may occur.
WARNINGS:
•
Disconnect the deflated cuff from the fill line connector when a procedure
is completed. Disconnecting a deflated cuff manages potential risks by:
(1) eliminating the risk of injury should a user unintentionally re-inflate a
deflated, unmonitored cuff and (2) providing confirmation that the cuff
deflation sequence has been implemented and completed. Failure to
comply may result in patient injury.
•
At the end of a procedure, as soon as the tourniquet pressure is released,
remove the cuff, sleeve and other underlying materials, as the slightest
impedance of venous return may lead to congestion and pooling of blood
in the operative field. Failure to comply may result in patient injury.
Follow guidelines developed by your institution regarding standards of practice
for tourniquet use. Failure to follow instructions could result in possible
medical complications.
Published tourniquet guidelines and training programs developed by the Association of
periOperative Registered Nurses (AORN) provide valuable references and resources to establish
and update guidelines.
The AORN
Recommended Practices for Use of The Pneumatic Tourniquet
is available online.
Originally published in the
AORN Journal
and revised in December 1998, it is included in the
AORN
2000 Standards, Recommended Practices, and Guidelines
.
AORN’s web address is www.aorn.org. The AORN mailing address is 2170 South Parker Rd, Suite
300, Denver, CO 80231.
Environment and Placement
WARNINGS:
♦
The SmartPump is designed for use in a non-sterile zone of a surgical operating room. The
SmartPump must be placed outside the patient zone, 6 feet [1.83 m] horizontally 8 feet
[2.5 m] vertically.
♦
For optimal use, the SmartPump, including the roll stand with pole, must be positioned and
installed according to the instructions herein.
♦
Use only hospital grade line cord connected to a hospital grade grounded receptacle.
♦
Ensure power cord is away from personnel traffic and areas of water or liquids.
♦
DO NOT place the SmartPump on an unstable cart, stand, or table. Recommend using the
Stryker Roll Stand with pole REF 5920-013-000.
♦
DO NOT use this equipment in the presence of a mixture consisting of flammable
anesthetic and air or oxygen or nitrous oxide.