General Use Procedure
23
Setting Timer and Target Pressure
To adjust the cuff inflation timer or the target pressure, follow the
instructions below. Newly set values will remain in effect until they are:
(1) changed, (2) the
Default Display
button is pressed after deflation
or, (3) the unit is turned off and restarted.
NOTE:
A quick
error
-
beep
sounds if an inappropriate button press
sequence is used. For example, if the (
+
) or (
-
) button is pressed
before pressing
Set/Save,
the unit produces an
error-beep
.
1. Time and Pressure values may be changed when the
SmartPump is in the ‘Ready’ mode or during active cuff inflation.
2. Press the
Set/Save
button on the side requiring change.
The corresponding Time and Pressure values will blink.
3.
To change the set Time,
use the increase (
+
) or decrease (
-
)
buttons. Each press adjusts the time in 5-minute increments (1-
minute increments in the 1 to 15 minute time range). The
numeric time display changes accordingly.
4. To change the set Pressure, use the increase (+) or decrease (-)
buttons. Each button press adjusts the pressure in 5-mmHg
increments, displayed numeric values change accordingly.
5. When the desired value is displayed, press the
Set/Save
button
to implement the new settings. A confirmation
double-beep
indicates the new values are accepted. The blinking display
returns to normal operation.
6. If a cuff is not inflated, the
Ready
indicator reappears in the
center of the display screen. If settings are changed while a cuff
is in use, the cuff Pressure and/or Time changes are real-time.
7. When using two cuffs, the cuffs are controlled separately.