Preparing for Use
19
Default Parameters
SmartPump factory preset default time and pressure settings are:
1.
Elapsed Procedure Time: One hour
(Displayed as:
1:00
)
2.
Target Cuff Pressure: 250 mmHg
(Displayed as:
250
)
Procedure Timer
The timer display shows elapsed procedure time.
The timer may be set to the desired cuff inflation time from a minimum
of one minute. Timer adjustment may be made in one-minute
increments from 1 to 15 minutes, and five minute increments from 15
to 240 minutes.
The Cuff time alarm is preset at the factory for one hour. This means
that, if unchanged, the time setting for the Cuff is one hour. The
Timer’s
alert will sound, the
Alarm Indicator Mute
button will flash red
and the corresponding time display will flash when the preset total
elapsed inflated cuff time has been reached.
N
OTE
:
The SmartPump does
NOT
automatically deflate the cuff when
the target time is reached. It maintains inflation until the
Deflate
button
is pressed or the SmartPump is turned off.
Target Pressure
The default target pressure is preset to 250 mmHg at the factory. The
default pressure may be adjusted and set between an operating range
of 100 to 475 mmHg, in 5 mmHg increments.
Содержание SmartPump Rx
Страница 53: ...Specifications 53 ...
Страница 54: ...Specifications 54 ...
Страница 55: ...Specifications 55 ...