Specifications 47
7 Specifications*
Model:
5920-011-000 SmartPump Dual Channel Tourniquet Pump
Size:
Dimensions:
15.5 inch height x 10 inch width x 9 inch depth [39.4cm x 25.4cm x 22.9cm];
add 1.5 inch [3.8cm] for mounting pole at bottom;
1 inch [2.54cm] diameter roll stand pole mount
Weight:
Unpackaged:
12.0 lbs [5.45 kgs]
Packaged:
18.25 lbs [8.28 kgs]
Equipment Type:
Class I, Defibrillation Proof Type BF Applied Part
Approvals
CAN/CSA C22.2 No. 601.1
UL 60601-1
IEC 60601-1-2:2001 using IEC 61000-4-2, -3, -4, -5, -6, -8, -11;
EN61000-3-2, -3; EN55011 level A/ CISPR 11 A emissions
Power Sources:
AC Input:
100 - 240 VAC 50/60 Hz universal switching power supply
Automatic internal battery switchover for AC–free use or AC loss protection
Facility power:
120 VAC, 10A, hospital grade receptacle
AC Input:
Input voltage range:
100 – 240 VAC
Frequency:
50/60 Hz
AC fuses:
3.15A / 250V
Power requirements:
1.5 A @ 120 VAC
Internal DC Supply Module:
Output voltage:
15 VDC ± 1%, 150 mV pk-pk ripple
Output current:
4A maximum
Output power:
60 W maximum
Overload protection:
short circuit, 110% - 150% auto recovery
Over voltage protection:
115% - 130% of rated output voltage
Internal Battery:
Type:
nickel metal hydride (NiMH)
Nominal voltage:
12 VDC
Rated capacity:
2.6 Ahr
Over current protection:
integral PTC fuse
Over temperature sense:
integral thermistor
Charge time:
1 hour (80% charge), 6 hours (full charge)
Power conservation:
30 days (fully charged, then unplugged)
Display:
Type:
High resolution backlit custom LCD
Timer:
Elapsed time numeric
Pressure:
Real-time digital numeric with graphic pressure gauge
*Specifications are approximate and may vary from unit to unit or as a result of power supply fluctuations.