![Stryker SmartPump Rx Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/stryker/smartpump-rx/smartpump-rx_instructions-for-use-manual_1386029049.webp)
Specifications 49
Electromagnetic Compatibility Information
This equipment has been tested and found to comply with the limits for medical devices to IEC 60601-1-
2:2001. These limits are designed to provide reasonable protection against harmful interference in a typical
medical installation. This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to other devices in
the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to other devices, which can be determined by turning the
equipment off (with AC power cord unplugged) and then on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•
reorient or relocate the receiving device
•
increase the separation between the equipment
•
connect the equipment into an outlet on a circuit breaker different from that to which the other
device(s) are connected
•
consult the manufacturer or field service technician for help.
Only Stryker printer and shielded serial cable may be connected to the equipment’s serial port. Usage of
accessories or cables other than those sold by Stryker may result in increased RF emissions or decreased
RF immunity of the equipment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Dual Channel SmartPump is intended for use in the electromagnetic environment specified below. The customer
or the user of the Dual Channel SmartPump should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Dual Channel SmartPump uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
Complies
The Dual Channel SmartPump is suitable for use in all establishments other
than domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.