Service Code Summary
46
6 Service Code Summary
WARNING:
DO NOT service this equipment. If you require service,
contact your Stryker sales representative or call Stryker customer service
at 1-800-253-3210. Failure to comply may result in user/patient injury.
Service Alarms
Code
Description
CALIBRATION ERROR
30
Calibration test detects error.
REAL-TIME CLOCK STOPPED
31
Real-time clock has failed or stopped.
SERIAL INTERFACE ERROR
32
Serial interface fault.
FP NONVOLATILE MEMORY ERROR
33
The nonvolatile memory for the FP processor has
failed a memory test.
FP FILES CORRUPT
34
The nonvolatile memory storing the FP code has
failed a memory test.
A NONVOLATILE MEMORY ERROR
35
The nonvolatile memory for the A code has failed a
memory test.
B NONVOLATILE MEMORY ERROR
36
The nonvolatile memory for the B code has failed a
memory test.
HARDWARE FAULT
37
An internal hardware fault was detected.
POWER SUPPLY FAULT
38
Internal power supply circuit fault was detected.
BLADDER 1 LEAK
40
Bladder 1 leaks or is not operating correctly.
BLADDER 2 LEAK
41
Bladder 2 leaks or is not operating correctly.
VALVE OR AIR PUMP FAULT
42
Fault detected with the operation of the air valves or
the air pump.