Cleaning and Maintenance
40
Periodic Maintenance
The SmartPump should be checked annually for proper operation.
Required hardware and materials:
♦
Calibrated manometer (0 to 500 mmHg range)
♦
10 foot fill line
♦
15 inch to 24 inch Stryker Color Cuff for test purposes
♦
Y-connector to connect manometer to fill line when
measuring pressure
♦
Soft clean cloth and cleaning agent (see
Cleaning
Recommendations
).
N
OTE
:
Test both the Cuff 1 and Cuff 2 sides of the SmartPump.
Complete one side before attempting the other side.
Perform the following steps annually. These guidelines are not
intended to replace inspection and checks that are independently
developed and adopted by the end user’s institution.
1.
Examine the enclosure, all the connectors, the AC power cord
and the fill line for damage, such as cracks, abrasion, or a
damaged ‘O’ ring on the fill line connector. DO NOT use the
equipment if damage is apparent.
2.
Clean the SmartPump. See the
Cleaning Recommendations
.
3.
Plug the SmartPump into AC power. Verify that the green AC
power light on the front panel is lit.
4.
Verify the internal battery charge light is either blinking yellow
(fast charge mode), solid yellow (slow charge mode) or is
solid green (trickle charge mode).
5.
Allow the SmartPump’s internal battery to charge fully. A
complete charge should take no more than six hours. Charge
time depends on how much the battery was discharged.
When fully charged, the internal battery charge light will
illuminate green.
6.
Press and release the
ON/OFF
(standby) button on the front
panel. During the startup sequence, the SmartPump’s
software version is displayed at the lower left of the LCD.
Record the software version, along with the serial number, on
the checklist form.