Cleaning and Maintenance
41
7.
The SmartPump should now be at the ‘ready’ mode, with the
default time and default pressure indicated. If the default
values are not: 1:00 hour (60 minutes if using the Minutes
time format) and 250 mmHg pressure, use the
Set/Save
button to adjust the settings, then proceed.
8.
Remove AC power by unplugging the AC power cord. The two
LEDs on the front panel (AC power and internal battery
charge) should turn off. The SmartPump should remain
operating, using the internal battery to provide power. The
Alarm Indicator Mute
button should flash red and the
n o A C
message should appear on the lower left of the LCD. Press
the
Alarm Indicator Mute
button to cause the
Alarm
Indicator Mute
button to remain green.
9.
Turn the unit off and then back on using the front panel
ON/OFF
(standby) button. The two LEDs on the front panel
(AC power and battery charge) should be off. The unit will be
operating using the internal battery for power. The
Alarm
Indicator Mute
button should flash red and the
n o A C
message should appear on the lower left of the LCD. Press
the
Alarm Indicator Mute
button to cause the
Alarm
Indicator Mute
button to stay solid green color.
10.
Change the Time setting to 15 minutes. Press
Set/Save
button to save this new time setting.
11.
Connect a 10-foot fill-line to the SmartPump. Connect a Y-
connector to the end of the fill line. Connect a 15 inch to 24
inch size cuff to one branch of the Y-connector. Connect the
manometer to the other branch of the Y-connector.
12.
Turn the manometer on and zero the pressure value of the
manometer. See the instructions for use supplied with the
manometer.
13.
Allow the cuff to lie flat on a smooth surface. Ensure the
fastener does not restrict the cuff from filling to a ‘tube’
shape during the testing.
14.
Confirm the pressure is set to 250 mmHg. Press the
Inflate
button
and allow the pressure to stabilize for 5 seconds. The
tube shape cuff may need a few seconds to stabilize its
volume.
15.
Record the displayed pressure and manometer pressure on
the checklist.