15 Electromagnetic Compatibility
Guidance and manufacturer's declaration - electromagnetic emissions
The Neptune 3 Rover REF 0703-001-000 is intended for use in the electromagnetic environment specified below. The customer or the user of the
Neptune 3 Rover REF 0703-001-000 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Neptune 3 Rover REF 0703-001-000 uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The Neptune 3 Rover REF 0703-001-000 is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for domestic purposes,
provided the following warning is heeded:
Harmonic emissions
IEC 61000-3-2
Class A
WARNING:
This equipment/system is intended for use by healthcare professionals only.
This equipment/system may cause radio interference or may disrupt the operation of
nearby equipment. It may be necessary to take mitigation measures, such as reorienting or
relocating the Neptune 3 Rover REF 0703-001-000 or shielding the location.
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacturer's declaration - electromagnetic immunity
The Neptune 3 Rover REF 0703-001-000 is intended for use in the electromagnetic environment specified below. The customer or the user of the
Neptune 3 Rover REF 0703-001-000 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1kV for input/output
lines
± 2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Neptune 3 Rover REF 0703-001-000
requires continued operation during power mains interruptions, it
is recommended that the Neptune 3 Rover REF 0703-001-000 be
powered from an uninterruptible power supply or a battery.
Power frequency (60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristics of
a typical location in a typical commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.